Hello, welcome to General Measure's channel. In this video, we will show you how to troubleshoot an OFL error on a checkweigher.
During operation, an OFL error indicates that the weight exceeds the weighing range or the millivolt input is out of range. There are usually four possible causes. We will use the CW-3K checkweigher as an example.
The value entered does not match the used weight during static calibration.
The product weight exceeds the configured maximum capacity.
Incorrect load cell wiring.
A faulty load cell or main control board.
If OFL appears, stop the checkweigher and remove all items from the weighing platform. Then go to Settings → Static Calibration and check whether the zero-point millivolts is within the normal range of 3.7–4.6 mV.
If the zero-point signal is within the normal range, recalibrate the scale.
Tap the empty-platform calibration button, place the calibration weight on the platform, enter its actual weight, and then tap the weight calibration button.
If the zero-point signal is outside the normal range, continue troubleshooting under Cause 3 or 4.
In addition to static calibration issues, a product weight that exceeds the configured maximum capacity will also trigger an OFL alarm.
Follow these steps to check the maximum capacity setting:
On the main screen, tap Settings and open Operating Parameters. Tap Next Page until you reach page 4, then check the current maximum capacity setting.
If the product weight is within the configured capacity but the OFL alarm persists, check the load cell wiring and related hardware.
The load cell is installed and its wiring is checked before the equipment leaves the factory.
If the load cell has been replaced on site, refer to the signal definitions in the supplied manual. Make sure each wire is connected to the corresponding marking printed beside the load cell connector on the main control board.
If the zero-point signal remains outside the normal range of 3.7–4.6 mV after ruling out wiring errors and the other causes above, further checks of the load cell and main control board are required.
You will need a multimeter and a flat-head screwdriver. Follow these steps:
① Disconnect the SIG+ and SIG− signal wires
First, use a flat-head screwdriver to disconnect the SIG+ and SIG− signal wires from the main control board to prevent external influences from affecting the measurements.
② Measure the excitation voltage across EX+ and EX−
Then, set the multimeter to DC voltage and check whether the excitation voltage across EX+ and EX− is 5 V DC ±10%.
Within range: The board’s excitation voltage is normal. Continue by checking the load cell output signal.
Outside 5 V DC ±10%: The main control board is faulty and must be replaced. [Fault indication]
③ Measure the signal voltage across SIG+ and SIG−
If the excitation voltage is normal, set the multimeter to DC millivolts and measure the output voltage across the disconnected load cell SIG+ and SIG− wires. Check whether it is within 0–10 mV.
Within range: The signal voltage is normal. Continue by checking the load cell resistance.
Outside 0–10 mV: The load cell is faulty and must be replaced. [Fault indication]
④ Measure the load cell resistance
Disconnect the load cell connector from the main control board. Then use the multimeter to measure the following resistances:
Across EX+ and EX−: 382 Ω
Across SIG+ and SIG−: 350 Ω
Between EX+ and SIG+: 280.7 Ω
Between EX− and SIG−: 280.7 Ω
Under normal conditions, the resistance values should follow this relationship:
EX±(400±20Ω)>SIG±(350±5Ω)>EX+SIG+≈EX-SIG-(280±2Ω)
The resistance between EX+ and SIG+ should be approximately equal to that between EX− and SIG−.
If these two values differ significantly, or the resistance values do not follow the relationship above,
The load cell bridge is unbalanced and the load cell must be replaced.
If all the tests above are normal, the main control board is considered faulty and should be replaced.
Finally, if several units of the same model are available on site, you can swap the main control board with one from another unit to quickly identify the faulty component.